Defibrilator Bifazic HeartStart XL
Pret: la cerere
Producator: PHILIPS - OLANDA
Produs publicat de: MOBILIER SI APARATURA MEDICALA -- Timisoara (Timis) , telefon: 0745316171 / 0745070413
Producator: PHILIPS - OLANDA
Produs publicat de: MOBILIER SI APARATURA MEDICALA -- Timisoara (Timis) , telefon: 0745316171 / 0745070413
M4735A HeartStart XL Defibrillator/Monitor
Specificatii
Defibrillator
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted
as a function of patient impedance.
Shock Delivery: Via multifunction defib electrode pads, or paddles.
Charge Time: Less than 3 seconds to 200 Joules with a new, fully charged
M3516A SLA battery pack at 25oC.
Less than 15 seconds to 200 Joules when powered by AC with no battery installed.
Patient Impedance Range: 25 to 180 Ohms.
Manual Mode
Manual Output Energy (Delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150,
200 Joules.
Controls: Manual/AED On/Energyknob, Charge/Disarm, Shock,
ECG Lead Select, SpO2 On/Off, SpO2 Alarms, HR Alarms, Sync On/Off,
Pacer, Pacer Start/Stop, Pacer Rate, Pacer Current, Pacer Mode, ECG Gain,
Volume, Strip, Summary, Mark.
Indicators: EL display for ECG waveform and text prompts, Audio alerts,
QRS Beeper, Charging Tones (for sync and asynchronous modes),
AC Power LED, Battery Charging LED, Sync LED, Pacer LED.
Armed Indicators: Charge done tone and available energy indicated on
display.
Energy Selection: Front panel rotary knob.
Charge Control: Front Panel "2" key or button on paddles.
Table 7-1 Delivered Energy (J) vs. Load Impedance
Selected
Energy
(J)
Load Impedance (ohms)
Accuracy
25 50 100 125 150
2 1.9 2.0 2.1 2.1 2.1 ± 1 J
3 2.8 3.0 3.2 3.2 3.1 ± 2 J
5 4.7 5.0 5.2 5.4 5.2 ± 2 J
7 6.6 7.0 7.3 7.5 7.3 ± 2 J
10 9.3 10.0 10.4 10.7 10.4 ± 2 J
20 18.6 20.0 20.8 21.4 20.8 ± 4 J
30 27.9 30.0 31.2 32.1 31.2 ± 4 J
50 46.7 50.0 52.3 53.5 52.1 ±15%
70 65.4 70.0 73.1 75.0 72.9 ±15%
100 93.5 100.0 104.7 107.2 104.4 ±15%
150 140.3 150.0 156.8 161.0 156.5 ±15%
200 187.0 200.0 209.3 214.6 208.6 ±15%
M4735A Service Manual 7-3
Specifications
7
Shock Control: Front Panel "3" key or buttons on paddles.
Synchronizer: SYNC message appears on the monitor and is annotated on
the printer (if printing while in Sync Mode). An audible beep sounds with
each detected R-wave, while a tick mark on the monitor and printed strip indicate
the discharge points. Synchronizer delay is less than 60 msec from peak
of R-wave to peak current of the defibrillation discharge.
AED Mode
AED Energy Profile: Fixed Energy (150 Joules).
AED Shock Series: 1, 2, 3, or 4 shocks per series.
Shock Series Timer: off, 30, 60, 90, 120, 150, 180, or 210 seconds.
Text and Voice Prompts: Extensive text and audible messages guide user
through protocol.
AED Controls: On, Off, Pause/Resume, Analyze/Stop Analysis, Shock,
Lead Select, SpO2 On/Off, SpO2 Alarms, HR Alarms, ECG Gain, Volume,
Strip, Summary, Mark.
Indicators: EL display for ECG waveform and text prompts, Audio alerts,
Voice prompts, QRS Beeper, Charging Tone, Charge Done Tone, Printer,
AC Power LED, Battery Charging LED.
Armed Indicators: Charge Done Tone, Available Energy indicated on display,
Voice Message.
Patient Analysis: Per protocol, evaluates patient ECG and signal quality to
determine if a shock is appropriate and evaluates connection impedance for
proper defibrillation pad contact.
Shockable Rhythms: Ventricular fibrillation with amplitude 100 uV and
wide complex ventricular tachycardia with rates greater than 150 bpm.
Sensitivity and Specificity: Meets AAMI guidelines.
ECG Monitoring
Inputs: Single channel ECG may be viewed on display and printed. Pads
ECG is obtained through two multifunction defibrillation electrode pads.
Lead I, II, or III are obtained through the 3-lead ECG cable and separate
monitoring electrodes. With a 5-lead cable, leads aVR, aVL, aVF, and any one
of the V (1-6) leads can also be obtained.
Lead Fault: LEADS OFF message and dashed line appear on the display if
an electrode or lead wire becomes disconnected while a lead is selected.
Paddle Fault: NO PADDLES CONNECTED message and dashed line
appear on the display in Manual Mode if paddles become disconnected.
Pad Fault: PADS OFF message and dashed line appear on the display if a pad
becomes disconnected.
Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an
accuracy of ±10%.
Specifications
7-4 Specifications
Heart Rate Alarms: Configurable pairs of low and high heart rate alarm
limits: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Hands Free Defibrillation Patient Cable Length: 7 ft. (2.13 m).
ECG Cable Length: 12 ft. (3.7 m).
Common Mode Rejection: Greater than 90 dB measured per AAMI standard
for cardiac monitors (EC 13).
ECG Size: 2.5, 5, 10, 20, 40 mm/mV.
Frequency Response:
AC Line Filter: 60 Hz/50Hz.
Pads ECG for Display: Monitor (.15-40 Hz) or EMS (1-30 Hz).
Pads ECG for Printer: Monitor (.15-40 Hz) or EMS 1-30 Hz).
Leads ECG for Display: Monitor (.15-40 Hz) or EMS (1-30 Hz).
Leads ECG for Printer: Diagnostic (.05-150 Hz) or EMS (1-30 Hz) or
Monitor (.15-40 Hz).
Patient Isolation (defibrillation proof):
ECG: Type CF
SpO2: Type CF
Defib: Type BF - External paddles/pads
Type CF - Internal paddles
NOTE The HeartStart XL supports two types of Displays, EL or LCD. See
"Identifying the Display Type" on page 5-7 for information.
Display (EL)
Size: 115 mm x 86 mm.
Type: EL - Electroluminescent.
Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
Display (LCD)
Size: 111.4 mm x 83.5 mm.
Type: LCD - TFT Color Liquid Crystal Display
Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
M4735A Service Manual 7-5
Specifications
7
Battery
Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid (SLA).
Dimensions: 2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm (W) x 182 mm (D).
Weight: 1.4 lb. (0.65 kg).
Charge Time:
ô€º Approximately 3 hours to 90% charge, indicated by LED on front
panel.
ô€º Approximately 15 hours total to 100%.
Capacity: With a new, fully charged battery at room temperature (25oC):
ô€º 100 minutes ECG monitoring only, or
ô€º 50 full-energy (200 J) discharges, or
ô€º 75 minutes ECG monitoring while pacing (100 mA, 100 ppm).
Battery Indicators: LOW BATTERY message appears on display when at
least 10 minutes of monitoring time and 5 maximum-energy discharges
remain (with a new battery at room temperature - 25oC).
Battery Storage: Should not be stored above 40oC for extended periods of
time.
Unit can be operated using only AC power, with no battery installed.
Thermal Array Printer
Continuous Real Time Strip: User starts and stops the strip. The Print Strip
prints the selected ECG lead with the following data:
HEADER 1: Date, Time, Heart Rate, the SpO2 Value (if available), and
the text "Delayed" if printing has been configured for Delayed Mode.
Prints every 12 seconds.
HEADER 2: Current mode (AED/Manual), Lead, Gain, filter setting, the
text "Sync" (if Sync has been enabled), and Pacer Settings (consisting of
the Pacer Mode, Rate, and Current, if currently pacing the patient). Prints
every 12 seconds, with Header 1.
HEADER 3: Changes in Mode, Gain, Lead, Sync, and Pacer Settings.
FOOTER: Drug Annotations, HR/SpO2 limits on a Limit Alarm, the
Results of Analysis in AED Mode (No Shock Advised, Shock Advised, or
Cannot Analyze), Charging to xxxJ, Shock Delivered, No Shock Delivered,
Disarm, Battery Low.
SYMBOLS: Mark Triangle (for presses of the Mark key), an Alarm Bell
(Alarm Limit Violations), Lightning Bolt (Shock Delivered; followed by
b for biphasic), Vertical stripe Boundaries/Pacer/Sync Tick Marks.
Specifications
7-6 Specifications
Event Printing: Mark key automatically documents ECG and events during
defibrillation episodes. The Mark key can annotate the event with one of the
following labels: Epinephrine (Adrenaline), Atropine, Lidocaine, and Other.
Auto Printing: The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.
Delayed Printing: The printer can be configured to run real time or with a six
second delay.
Reports: The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log, Battery Capacity Test, Shift/System Check.
Speed: 25 mm/s with an accuracy of ±5%.
Amplitude Accuracy: ±10% or ±50 uV, whichever is greater.
Paper Size: 50 mm by 30 m (100 ft.).
Noninvasive Pacing
Waveform: Monophasic Truncated Exponential.
Current Pulse Amplitude:
Range: 10 mA to 200 mA (5 mA resolution)
Accuracy: 10 mA - 50 mA ±5 mA
50 mA - 200 mA ±10%.
Pulse Width: 20 ms with accuracy +0, -5 ms.
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy ±1.5%.
Modes: Demand or Fixed Rate.
Refractory Period: 80 ppm: 340 msec ±10%
80 ppm: 240 msec ±10%
SpO2 Pulse Oximetry
Range:
SpO2 - 0 to 100%.
Pulse Rate - 30 to 300 bpm.
Accuracy with:
M1191A sensor - 1 standard deviation 70% to 100%, ± 2.5%
M1192A sensor - 1 standard deviation 70% to 100%, ± 2.5%
M1194A sensor - 1 standard deviation 70% to 100%, ± 4.0%
M1131A sensor - 1 standard deviation 70% to 100%, ± 3.0%
M1903B sensor - 1 standard deviation 70% to 100%, ± 3.0%
M1904B sensor - 1 standard deviation 70% to 100%, ± 3.0%
Pulse Rate Accuracy: 2% or 1 bpm (whichever is greater).
Wavelength Range: 500 to 1000 nm.
M4735A Service Manual 7-7
Specifications
7
Emitted Light Energy: to 15 mW.
DisplayInterval: 60 seconds.
Resolution: 1%.
Sp02 Alarm Limits: Three preset low alarm limits: 90%, 85%, and 80%.
INOP Alerts: Triggered by disconnected sensor, noisy signal, light interference
or low signal (non-pulsatile).
Alarm delay: 10 seconds.
Event Storage
Internal Event Summary:
The internal Event Summary stores up to 300 events and up to 50 waveforms.
Events can be marked with a Mark symbol and, if configured for drug annotation,
the following labels can be added: Epinephrine (Adrenaline in U.K. and
Australia), Atropine, Lidocaine, or Other.
The Summary key on the front panel is used to print the internal Event Summary.
Data Card Event Summary:
The Data Card stores continuous ECG waveforms and events.
General
Dimensions: 19.0 cm (H) x 37.6 cm (W) x 34.6 cm (L)
7.5" x 14.8" x 13.7"
Weight: Standard Configuration weighs 6.5 kg (14.3 lbs.) including battery,
full roll of paper, and external paddles.
Specificatii
Defibrillator
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted
as a function of patient impedance.
Shock Delivery: Via multifunction defib electrode pads, or paddles.
Charge Time: Less than 3 seconds to 200 Joules with a new, fully charged
M3516A SLA battery pack at 25oC.
Less than 15 seconds to 200 Joules when powered by AC with no battery installed.
Patient Impedance Range: 25 to 180 Ohms.
Manual Mode
Manual Output Energy (Delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150,
200 Joules.
Controls: Manual/AED On/Energyknob, Charge/Disarm, Shock,
ECG Lead Select, SpO2 On/Off, SpO2 Alarms, HR Alarms, Sync On/Off,
Pacer, Pacer Start/Stop, Pacer Rate, Pacer Current, Pacer Mode, ECG Gain,
Volume, Strip, Summary, Mark.
Indicators: EL display for ECG waveform and text prompts, Audio alerts,
QRS Beeper, Charging Tones (for sync and asynchronous modes),
AC Power LED, Battery Charging LED, Sync LED, Pacer LED.
Armed Indicators: Charge done tone and available energy indicated on
display.
Energy Selection: Front panel rotary knob.
Charge Control: Front Panel "2" key or button on paddles.
Table 7-1 Delivered Energy (J) vs. Load Impedance
Selected
Energy
(J)
Load Impedance (ohms)
Accuracy
25 50 100 125 150
2 1.9 2.0 2.1 2.1 2.1 ± 1 J
3 2.8 3.0 3.2 3.2 3.1 ± 2 J
5 4.7 5.0 5.2 5.4 5.2 ± 2 J
7 6.6 7.0 7.3 7.5 7.3 ± 2 J
10 9.3 10.0 10.4 10.7 10.4 ± 2 J
20 18.6 20.0 20.8 21.4 20.8 ± 4 J
30 27.9 30.0 31.2 32.1 31.2 ± 4 J
50 46.7 50.0 52.3 53.5 52.1 ±15%
70 65.4 70.0 73.1 75.0 72.9 ±15%
100 93.5 100.0 104.7 107.2 104.4 ±15%
150 140.3 150.0 156.8 161.0 156.5 ±15%
200 187.0 200.0 209.3 214.6 208.6 ±15%
M4735A Service Manual 7-3
Specifications
7
Shock Control: Front Panel "3" key or buttons on paddles.
Synchronizer: SYNC message appears on the monitor and is annotated on
the printer (if printing while in Sync Mode). An audible beep sounds with
each detected R-wave, while a tick mark on the monitor and printed strip indicate
the discharge points. Synchronizer delay is less than 60 msec from peak
of R-wave to peak current of the defibrillation discharge.
AED Mode
AED Energy Profile: Fixed Energy (150 Joules).
AED Shock Series: 1, 2, 3, or 4 shocks per series.
Shock Series Timer: off, 30, 60, 90, 120, 150, 180, or 210 seconds.
Text and Voice Prompts: Extensive text and audible messages guide user
through protocol.
AED Controls: On, Off, Pause/Resume, Analyze/Stop Analysis, Shock,
Lead Select, SpO2 On/Off, SpO2 Alarms, HR Alarms, ECG Gain, Volume,
Strip, Summary, Mark.
Indicators: EL display for ECG waveform and text prompts, Audio alerts,
Voice prompts, QRS Beeper, Charging Tone, Charge Done Tone, Printer,
AC Power LED, Battery Charging LED.
Armed Indicators: Charge Done Tone, Available Energy indicated on display,
Voice Message.
Patient Analysis: Per protocol, evaluates patient ECG and signal quality to
determine if a shock is appropriate and evaluates connection impedance for
proper defibrillation pad contact.
Shockable Rhythms: Ventricular fibrillation with amplitude 100 uV and
wide complex ventricular tachycardia with rates greater than 150 bpm.
Sensitivity and Specificity: Meets AAMI guidelines.
ECG Monitoring
Inputs: Single channel ECG may be viewed on display and printed. Pads
ECG is obtained through two multifunction defibrillation electrode pads.
Lead I, II, or III are obtained through the 3-lead ECG cable and separate
monitoring electrodes. With a 5-lead cable, leads aVR, aVL, aVF, and any one
of the V (1-6) leads can also be obtained.
Lead Fault: LEADS OFF message and dashed line appear on the display if
an electrode or lead wire becomes disconnected while a lead is selected.
Paddle Fault: NO PADDLES CONNECTED message and dashed line
appear on the display in Manual Mode if paddles become disconnected.
Pad Fault: PADS OFF message and dashed line appear on the display if a pad
becomes disconnected.
Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an
accuracy of ±10%.
Specifications
7-4 Specifications
Heart Rate Alarms: Configurable pairs of low and high heart rate alarm
limits: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Hands Free Defibrillation Patient Cable Length: 7 ft. (2.13 m).
ECG Cable Length: 12 ft. (3.7 m).
Common Mode Rejection: Greater than 90 dB measured per AAMI standard
for cardiac monitors (EC 13).
ECG Size: 2.5, 5, 10, 20, 40 mm/mV.
Frequency Response:
AC Line Filter: 60 Hz/50Hz.
Pads ECG for Display: Monitor (.15-40 Hz) or EMS (1-30 Hz).
Pads ECG for Printer: Monitor (.15-40 Hz) or EMS 1-30 Hz).
Leads ECG for Display: Monitor (.15-40 Hz) or EMS (1-30 Hz).
Leads ECG for Printer: Diagnostic (.05-150 Hz) or EMS (1-30 Hz) or
Monitor (.15-40 Hz).
Patient Isolation (defibrillation proof):
ECG: Type CF
SpO2: Type CF
Defib: Type BF - External paddles/pads
Type CF - Internal paddles
NOTE The HeartStart XL supports two types of Displays, EL or LCD. See
"Identifying the Display Type" on page 5-7 for information.
Display (EL)
Size: 115 mm x 86 mm.
Type: EL - Electroluminescent.
Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
Display (LCD)
Size: 111.4 mm x 83.5 mm.
Type: LCD - TFT Color Liquid Crystal Display
Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
M4735A Service Manual 7-5
Specifications
7
Battery
Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid (SLA).
Dimensions: 2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm (W) x 182 mm (D).
Weight: 1.4 lb. (0.65 kg).
Charge Time:
ô€º Approximately 3 hours to 90% charge, indicated by LED on front
panel.
ô€º Approximately 15 hours total to 100%.
Capacity: With a new, fully charged battery at room temperature (25oC):
ô€º 100 minutes ECG monitoring only, or
ô€º 50 full-energy (200 J) discharges, or
ô€º 75 minutes ECG monitoring while pacing (100 mA, 100 ppm).
Battery Indicators: LOW BATTERY message appears on display when at
least 10 minutes of monitoring time and 5 maximum-energy discharges
remain (with a new battery at room temperature - 25oC).
Battery Storage: Should not be stored above 40oC for extended periods of
time.
Unit can be operated using only AC power, with no battery installed.
Thermal Array Printer
Continuous Real Time Strip: User starts and stops the strip. The Print Strip
prints the selected ECG lead with the following data:
HEADER 1: Date, Time, Heart Rate, the SpO2 Value (if available), and
the text "Delayed" if printing has been configured for Delayed Mode.
Prints every 12 seconds.
HEADER 2: Current mode (AED/Manual), Lead, Gain, filter setting, the
text "Sync" (if Sync has been enabled), and Pacer Settings (consisting of
the Pacer Mode, Rate, and Current, if currently pacing the patient). Prints
every 12 seconds, with Header 1.
HEADER 3: Changes in Mode, Gain, Lead, Sync, and Pacer Settings.
FOOTER: Drug Annotations, HR/SpO2 limits on a Limit Alarm, the
Results of Analysis in AED Mode (No Shock Advised, Shock Advised, or
Cannot Analyze), Charging to xxxJ, Shock Delivered, No Shock Delivered,
Disarm, Battery Low.
SYMBOLS: Mark Triangle (for presses of the Mark key), an Alarm Bell
(Alarm Limit Violations), Lightning Bolt (Shock Delivered; followed by
b for biphasic), Vertical stripe Boundaries/Pacer/Sync Tick Marks.
Specifications
7-6 Specifications
Event Printing: Mark key automatically documents ECG and events during
defibrillation episodes. The Mark key can annotate the event with one of the
following labels: Epinephrine (Adrenaline), Atropine, Lidocaine, and Other.
Auto Printing: The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.
Delayed Printing: The printer can be configured to run real time or with a six
second delay.
Reports: The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log, Battery Capacity Test, Shift/System Check.
Speed: 25 mm/s with an accuracy of ±5%.
Amplitude Accuracy: ±10% or ±50 uV, whichever is greater.
Paper Size: 50 mm by 30 m (100 ft.).
Noninvasive Pacing
Waveform: Monophasic Truncated Exponential.
Current Pulse Amplitude:
Range: 10 mA to 200 mA (5 mA resolution)
Accuracy: 10 mA - 50 mA ±5 mA
50 mA - 200 mA ±10%.
Pulse Width: 20 ms with accuracy +0, -5 ms.
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy ±1.5%.
Modes: Demand or Fixed Rate.
Refractory Period: 80 ppm: 340 msec ±10%
80 ppm: 240 msec ±10%
SpO2 Pulse Oximetry
Range:
SpO2 - 0 to 100%.
Pulse Rate - 30 to 300 bpm.
Accuracy with:
M1191A sensor - 1 standard deviation 70% to 100%, ± 2.5%
M1192A sensor - 1 standard deviation 70% to 100%, ± 2.5%
M1194A sensor - 1 standard deviation 70% to 100%, ± 4.0%
M1131A sensor - 1 standard deviation 70% to 100%, ± 3.0%
M1903B sensor - 1 standard deviation 70% to 100%, ± 3.0%
M1904B sensor - 1 standard deviation 70% to 100%, ± 3.0%
Pulse Rate Accuracy: 2% or 1 bpm (whichever is greater).
Wavelength Range: 500 to 1000 nm.
M4735A Service Manual 7-7
Specifications
7
Emitted Light Energy: to 15 mW.
DisplayInterval: 60 seconds.
Resolution: 1%.
Sp02 Alarm Limits: Three preset low alarm limits: 90%, 85%, and 80%.
INOP Alerts: Triggered by disconnected sensor, noisy signal, light interference
or low signal (non-pulsatile).
Alarm delay: 10 seconds.
Event Storage
Internal Event Summary:
The internal Event Summary stores up to 300 events and up to 50 waveforms.
Events can be marked with a Mark symbol and, if configured for drug annotation,
the following labels can be added: Epinephrine (Adrenaline in U.K. and
Australia), Atropine, Lidocaine, or Other.
The Summary key on the front panel is used to print the internal Event Summary.
Data Card Event Summary:
The Data Card stores continuous ECG waveforms and events.
General
Dimensions: 19.0 cm (H) x 37.6 cm (W) x 34.6 cm (L)
7.5" x 14.8" x 13.7"
Weight: Standard Configuration weighs 6.5 kg (14.3 lbs.) including battery,
full roll of paper, and external paddles.